Chapter 5 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 53

7. Select menu item [4] Point 2 and press the

Set button. The HM40 now shows the
currently measured value of the selected
quantity. Set the reference value using the
arrow buttons and press the OK button.

The correction to the measurement at point 2
is now shown in the text for the menu item
[4].

8. Select menu item [5] Note to edit the

calibration info text that is stored in the probe.
Edit the text using the select button and arrow
keys. When done, select the OK character in
the bottom right corner to save the changed
text.

To exit without saving, press the Cancel
button.

9. Select menu item [6] Apply to save the

calibration in the probe. Check the applied
corrections in the confirmation screen and
press the Apply button.

To exit without applying the correction, press
the Cancel button.