SETUP MODE

P3.0 View Slope & Offset

1. Press MODE to select probe type as pH.

2. Press SET for 2 seconds to enter setup.

3. Press until the icon “ELE” appears in the middle

display and P3.0 appears in the lower display.

4. Press MN/MX/AV to enter P3.1, the LCD displays one of

four available slope values; P3.1, P3.2, P3.3, P3.4. If
the value is less than 75% or more than 115%, change
the probe immediately.

5. Press MN/MX/AV to enter P3.2, P3.3, and P3.4.

Note...
The solution range definition differs between NIST and
Custom buffers.