Boonton 4540 Peak Power Meter Quick Start User Manual
Page 41

4540 Power Meter
Quick Start Guide
PN: 98406000A
© Copyright 2008 Boonton
Page 41
Warm-up (optimal)
While a minimum warm-up and stabilization period of 3 mins is required, for best
possible calibration and measurement accuracy we recommend connecting the
sensor to the 4540 Peak Power meter with power switched on 15 mins before the
calibration is initialized. Warm up and stabilization period is only required for Peak
Power Sensors.
Zero Calibration
Term commonly used when sensor calibration at 0 mW