Performing Calibration

1. Connect the sensor 1 to Channel 1 of the 4540.
2. Connect the sensor 2 to Channel 2 of the 4540 – if applicable.

Note:

To ensure accurate calibration of Peak power sensors it is strongly

recommended to allow a warm-up and stabilization period of at least 3 minutes,
after a sensor has been connected to a powered-on instrument. For best
possible calibration we recommend to extend this period to 15 minutes.


Calibrating Sensor 1
3. Connect sensor 1 to the Cal Out connector.
4. Press Cal Button to select the Calibration Menu
5. Display shows the cal menu with “CH1>Calibration” at the top.
6. Press AutoCal Start

Calibration of sensor
connected to CH1 is performed

7. Disconnect Sensor 1 from

Cal Out


On a 4542 dual channel
system with two sensors

Calibrating Sensor 2
8. Connect Sensor 2 to the

Cal Out

connector.

9. Press Cal Button again to

select the Calibration Menu

10. Display shows the cal menu with “CH2>Calibration” at the top.
11. Press AutoCal Start

Calibration of sensor connected to CH2 is performed

12. Disconnect Sensor 2 from Cal Out

Note:

AutoCal selects calibration points based on the dynamic range of the

connected sensor. AutoCal will extend the calibration range typically about 3dB of
the sensor specification.