Boonton 4540 Peak Power Meter Quick Start User Manual
Page 17

4540 Power Meter
Quick Start Guide
PN: 98406000A
© Copyright 2008 Boonton
Page 17
Performing Calibration
1. Connect the sensor 1 to Channel 1 of the 4540.
2. Connect the sensor 2 to Channel 2 of the 4540 – if applicable.
Note:
To ensure accurate calibration of Peak power sensors it is strongly
recommended to allow a warm-up and stabilization period of at least 3 minutes,
after a sensor has been connected to a powered-on instrument. For best
possible calibration we recommend to extend this period to 15 minutes.
Calibrating Sensor 1
3. Connect sensor 1 to the Cal Out connector.
4. Press Cal Button to select the Calibration Menu
5. Display shows the cal menu with “CH1>Calibration” at the top.
6. Press AutoCal Start
Calibration of sensor
connected to CH1 is performed
7. Disconnect Sensor 1 from
Cal Out
On a 4542 dual channel
system with two sensors
Calibrating Sensor 2
8. Connect Sensor 2 to the
Cal Out
connector.
9. Press Cal Button again to
select the Calibration Menu
10. Display shows the cal menu with “CH2>Calibration” at the top.
11. Press AutoCal Start
Calibration of sensor connected to CH2 is performed
12. Disconnect Sensor 2 from Cal Out
Note:
AutoCal selects calibration points based on the dynamic range of the
connected sensor. AutoCal will extend the calibration range typically about 3dB of
the sensor specification.