Boonton 4540 Peak Power Meter Quick Start User Manual
Page 13

4540 Power Meter
Quick Start Guide
PN: 98406000A
© Copyright 2008 Boonton
Page 13
Rear Panel Measurement and Calibrator
Connectors (Optional)
The 4540 can be ordered with measurement
channel(s) and/or the calibrator output mounted
on the rear panel. This is usually handy when
the instrument is integrated in systems like ATE.
Communication interfaces.
4540 comes standard with LAN (Ethernet), USB and GPIB
interface. All interfaces can be used to remote control the
instrument. When either GPIB, LAN or USB assume control of
the instrument, the keyboard entry is
automatically disabled. A remote sign will be
shown at the lower right side of the
screen during remote operation.
Pressing Esc / Stop / Local when the instrument is
in remote mode will return it to local mode
(the instrument is under keyboard control).
LAN
allows DHCP or fixed (IP / Subnet) setting mode.
USB
is compatible to V 1.1 and 2.0
GPIB
interface supports AH1, SH1, T6, LE0, SR1, RL1, PP0, DC1, DT1, C0, and
E1.
LAN, USB and GPIB ports are also used for 4540 firmware updates.