beautypg.com

3 functional specifications – AirSep MN134-1 B User Manual

Page 9

background image

MN134-1 rev B

02/14

4

1.3 Functional Specifications

Oxygen Concentration:

2-9 LPM: 92% +3.5/-3%

10 LPM: 90% +5.5/-3%

(Based on 70°F [21°C] at sea level)

Accuracy:

Flowmeter ± 10% or ± 200ml of indicated flow, whichever is
greater.

Response Time:

Allow 5 minutes to attain maximum oxygen concentration.

Positioning:

Operate the unit in an upright position, maintaining at least
12 inches (30.5cm) of open space on all sides for ventilation.

Dimensions:

27.5 in. high x 16.5 in. wide x 14.5 in. deep

(69.9 cm high x 41.9 cm wide x 36.8 cm deep)

Weight:

58 lb; shipping weight – 65 lb

(26.3 kg; shipping weight – 29.5 kg)

Electrical:

120 VAC, 60 Hz, 6.0 amps, 590 watts

220-240 VAC, 50 Hz, 3.0 amps, 590 watts

Two-prong polarized plug

Double-insulated cabinet

Alarms:

Power Failure

Low Oxygen Concentration (with optional Oxygen Monitor)
Low Pressure
High Pressure

High Temperature

Battery Test

Operating Temperature

41°F to 104°F (5°C to 40°C) for 120 VAC 60 Hz models

Range:

41°F to 95°F (5°C to 35°C) for 220-240 VAC 50 Hz models

Storage Temperature

-4°F to 140°F (-20°C to 60°C)

Range:

Humidity:

Up to 95% (non-condensing)