3 functional specifications – AirSep MN134-1 B User Manual
Page 9
MN134-1 rev B
02/14
4
1.3 Functional Specifications
Oxygen Concentration:
2-9 LPM: 92% +3.5/-3%
10 LPM: 90% +5.5/-3%
(Based on 70°F [21°C] at sea level)
Accuracy:
Flowmeter ± 10% or ± 200ml of indicated flow, whichever is
greater.
Response Time:
Allow 5 minutes to attain maximum oxygen concentration.
Positioning:
Operate the unit in an upright position, maintaining at least
12 inches (30.5cm) of open space on all sides for ventilation.
Dimensions:
27.5 in. high x 16.5 in. wide x 14.5 in. deep
(69.9 cm high x 41.9 cm wide x 36.8 cm deep)
Weight:
58 lb; shipping weight – 65 lb
(26.3 kg; shipping weight – 29.5 kg)
Electrical:
120 VAC, 60 Hz, 6.0 amps, 590 watts
220-240 VAC, 50 Hz, 3.0 amps, 590 watts
Two-prong polarized plug
Double-insulated cabinet
Alarms:
Power Failure
Low Oxygen Concentration (with optional Oxygen Monitor)
Low Pressure
High Pressure
High Temperature
Battery Test
Operating Temperature
41°F to 104°F (5°C to 40°C) for 120 VAC 60 Hz models
Range:
41°F to 95°F (5°C to 35°C) for 220-240 VAC 50 Hz models
Storage Temperature
-4°F to 140°F (-20°C to 60°C)
Range:
Humidity:
Up to 95% (non-condensing)