Eclipse User Manual — 15
Pulse Dose Mode and autoSat Features
Your Eclipse has a feature, called autoSAT
®
that delivers a
consistent-sized bolus of oxygen up to 40 breaths per minute (see
Max Breath Rate table below). As your breath rate increases, ie.
during ambulation, the autoSAT feature servo-controls the device,
delivering the set bolus size. autoSAT Technology is proprietary
to CAIRE and is utilized during Pulse Mode only.
Mas Breath Rate
Bolus
volume (+/- 15%)
Pulse
Dose
Setting
Bolus Size
(± 15%)
mL
AC Power Supply
& Power Cartridge
(Battery)
DC Power Supply
Max Breath Rate
Max Breath Rate
1.0
16
40
40
2.0
32
40
40
3.0
48
40
40
4.0
64
40
31
5.0
80
37
25
6.0
96
31
20
-
128
23
15
-
160
18
12
-
192
15
10
Step 5: Adjust the Flow Setting to the Prescribed
Level
Using the Flow Setting Buttons, adjust the flow setting to the
prescribed setting supplied by your physician.
Note: Bolus volume decreases as breath rate exceeds published range.
Note: The pulse dose setting (bolus) is not equal to LPM.
WARNING: PULSE DOSE MODE SETTINGS MUST BE DETER-
MINED FOR EACH PATIENT INDIvIDUALLY FOR THEIR NEEDS
AT REST, DURING EXERCISE AND WHEN TRAvELING.
WARNING: THE ECLIPSE MAY NOT BE ABLE TO DETECT ALL INSPIRA-
TORY EFFORTS IN PULSE MODE. IF THE ECLIPSE DOES NOT RESPOND
TO YOUR INSPIRATION EFFORTS, ASK YOUR HOME CARE PROvIDER
TO CHECK THE SENSITIvITY SETTING.
Note: To Equipment Provider: The following oxygen administration acces-
sories are recommended for use with the Oxygen Concentrator:
• Nasal Cannula with 7 feet (2.1 m) of tubing: PN 5408-SEQ
• For any additional recommended accessories, please see the Accessories
Catalog (PN ML-LOX0010) available on www.CAIREmedical.com.
WARNING: IT IS vERY IMPORTANT TO SELECT ONLY THE
PRESCRIBED LEvEL OF OXYGEN. DO NOT CHANGE THE
FLOW SELECTION UNLESS YOU HAvE BEEN DIRECTED TO
DO SO BY A LICENSED CLINICIAN.
Step 6: Begin Using Your Eclipse
Breathe normally through the nasal cannula.
Note: You will achieve longer operating time on the Battery if you operate
your Eclipse in the Pulse Mode (Physician approved mode).
Step 7: Universal Cart Operation
The Universal Cart is designed for use on pedestrian-type
surfaces.
Align the locater holes on the bottom of the device to the tabs on
the Universal Cart.
Align the large threaded screw to the
insert at the back of the Eclipse. Then
hand-tighten the knob on the cart until the
Universal Cart is tightly secured to the
device.
Depress the push-button on the Universal
Cart handle to adjust the height of the
handle.
Step 8: Power Off the Device
Press and hold the “ON/OFF” Button for two (2) seconds to
power OFF your Eclipse.
Store the Eclipse in a cool and dry location for a period no longer
than 1 month. The Eclipse must run for a minimum of 2 hours per
month to ensure proper function of the unit.
Note: DO NOT operate the Eclipse and ambulate while a humidifier is
attached. Remove the humidifier bottle before walking. DO NOT lay the
Eclipse down while attached to a humidifier bottle.