Setting the pediatric flowmeter – AirSep MN164-1 User Manual
Page 27

MN164-1 Rev -
08/09
23
Setting the Pediatric Flowmeter
When using the pediatric flowmeter, the unit will not reach proper concentration at the pediatric setting (less than 2 lpm) until
you bleed off a portion of the oxygen by opening the primary flowmeter (on the left side of the unit). Follow the procedure
below when using the pediatric flowmeter.
1. Follow the start-up instructions 1-7 as outlined on page 8 & 9.
2. Set the pediatric flowmeter to the prescribed flow.
3. Set the primary flowmeter to 2 lpm to bleed off excess product, and allow the unit to achieve maximum concentration.
The NewLife Intensity Oxygen Concentrator must be operated for
at least five minutes at 2 lpm before using the unit.
The NewLife Intensity is appropriate for usage by two patients,
provided the combined flow is a minimum of 2 lpm and does not
exceed the maximum capacity of the concentrator.
Figure 13
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN179-1 B (62 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)