beautypg.com

Agilent Technologies E1364A User Manual

Page 26

background image

380 OUTPUT @Switch;"CLOS (@100:115)"

!Open all NC contacts

390 FOR I=0 TO 15

400 !Closed-channel resistance (NC contacts)

410 OUTPUT @Switch;"OPEN (@"&VAL$(100+I)&")"

420 OUTPUT @Dmm;"TRIG SGL"

430 ENTER @Dmm;Rdg

440 PRINT "Channel "&VAL$(I)&" (closed): "&VAL$(Rdg)

450 !

460 !Open-channel resistance (NC contacts)

470 OUTPUT @Switch;"CLOS (@"&VAL$(100+I)&")"

480 OUTPUT @Dmm;"TRIG SGL"

490 ENTER @Dmm;Rdg

500 PRINT "Channel "&VAL$(I)&" (open): "&VAL$(Rdg)

510 PRINT

520 NEXT I

530 !

540 END

26 Verification Tests

Chapter 2