Performance verification – Agilent Technologies E1364A User Manual
Page 21
![background image](/manuals/40077/21/background.png)
Performance Verification
The procedures in this section are used to test the switch module’s electrical
performance using the specifications in Appendix A of the E1364A User’s
Manual as the performance standards. These tests are suitable for incoming
inspection, troubleshooting, and preventive maintenance.
Test Fixture
A test fixture is required for the following tests. It is recommended that you
order an extra terminal module (aka "terminal block") to use as a test
fixture, so that you do not have to re-wire the terminal module each time
these tests are performed. The terminal module part number is
E1364-80001.
Figures 2-1(a) and 2-1(b) show how the test fixture should be wired.
Perform the following steps to wire the test fixture:
–
Short all NO (Normally Open) lines together.
–
Short all NC (Normally Closed) lines together.
–
Short all C (Common) lines together.
Figure 2-1(a). Agilent E1364A Test Fixture Schematic
Chapter 2
Verification Tests 21