Agilent Technologies E1364A User Manual
Page 24

Normally Closed
Contacts
1. Hardware Connections
2. Equipment Setup
•
Set DMM to: 4-wire ohms, autorange.
•
Send
*RST
to the switch.
•
Send
CLOS (@100:115)
to the switch to open all NC contacts.
3. Closed-Channel Reading (channel 00)
•
Send
OPEN (@100)
to the switch to close channel 00.
•
Trigger the DMM and record the reading.
4. Open-Channel Reading (channel 00)
•
Send
CLOS (@100)
to the switch to open channel 00.
•
Trigger the DMM and verify that an open circuit is indicated
(>10
8
Ω
).
5. Closed- and Open-Channel Readings (channels 01-15)
•
Repeat steps 3 and 4 for channels 01 through 15. Use
CLOS (@ccnn)
and
OPEN (@ccnn)
, where
cc = switch card number (01-99, leading zero not necessary)
nn = channel number (00-15)
Figure 2-3. Closed-Channel Resistance (NC Channels)
24 Verification Tests
Chapter 2