MEASUREMENT LOOP VALIDATION
AirChip 3000 Functions
Description
► Simulator mode
Used to make the HF7 generate fixed values for the humidity,
temperature and calculated parameter. Can be configured,
enabled or disabled
DEVICE SAFEGUARDS
AirChip 3000 Functions
Description
► Device write protection
Used to protect the HF7 with a password to prevent
unauthorized digital access by a digital user. Can be configured,
enabled or disabled
PROCESS PROTECTION / PROTECTION OF OTHER DEVICES
AirChip 3000 Functions
Description
► Limit humidity output to 100 %RH
Used to prevent the humidity signal from exceeding
100 %RH when condensation forms on the sensor. Can be
enabled or disabled
► Out-of-limit value alarm
Used to specify the normal range for humidity, temperature and
the calculated parameter depending on the user application.
Can be configured, enabled or disabled
Out-of-limit values trigger a digital alarm which can be also be
seen on the optional display
► Bad sensor alarm
Built-in function. Cannot be disabled
A bad humidity or temperature sensor triggers a digital alarm
which can be also be seen on the optional display
► Fail safe mode
Used to specify a "safe" fixed value for humidity and for
temperature in the event of a sensor failure. Can be configured,
enabled or disabled
4.2 Factory default settings
Notes:
o Configuration of the HF7 by the user and access to its functions requires a PC with the ROTRONIC HW4
software (version 2.1.1 or higher) installed. Service cable AC3006 or AC3009 is used to connect the HF7
service connector to a USB port of the PC (see Maintenance > Service Cable).
Configurable Settings
Applicability
Factory default
Unit system (Metric or English)
HF72, HF73
As per ordering code
Analog signal type (4…20 mA or other)
HF73
As per ordering code
Psychrometric calculation
HF72, HF73
As per ordering code
Output 1 parameter, scale and unit
HF72, HF73
As per ordering code (%RH or DP)
Output 2 parameter, scale and unit
HF72, HF73
Temperature, unit as per ordering code
Functions
Applicability
Factory default