HygroMet MP Humidity Temperature
Meteorological Probe: User Guide
Instruction Manual
Document Type
Page 9 of 19
Document title
© 2009; Rotronic AG E-M-HM-V1_10
Out-of-limit value digital / display alarm
Disabled (HygroMet and probe)
Data recording (probe)
Enabled (loop mode – 10 min. interval)
Automatic humidity sensor test
Disabled (probe)
Humidity sensor drift compensation
Disabled (probe)
Fail safe mode
Disabled (probe)
Monitor probe alarms
Enabled
Loss of communication with probe
Disabled (HygroMet)
Simulator mode
Disabled (HygroMet and probe)
o
For a detailed description of all AirChip 3000 main functions see document E-T-AC3000-DF-V1
o
Instructions regarding the configuration of the HygroMet MP and probe as well as access to the functions
are provided in the following manuals:
E-M-HW4v2-F2-014
E-M-HW4v2-F2-001
E-M-HW4v2-Main (§ 6.5)
E-M-HW4v2-DR-001
E-M-HW4v2-A2-001
E-M-AC3000-CP
o
The factory default setting for dew / frost point calculation is frost point below freezing
4.4 Interaction between the HygroMet MP and probe functions
It is important to note that when used together, the HygroMet MP transmitter and HC2 probe (HygroClip 2)
constitute a 2-component system. Each system component has its own microprocessor, firmware and
functions. Some of these functions are unique to each system component. Other functions are found in both
components.
The functions and settings of the HygroMet MP (HM) and HygroClip 2 probe (HC2) operate together as
indicated below:
Function / Setting
HM
HC2
Notes
Device protection
X
X
Individual to the HygroMet and HC2 probe
RS-485 address
X
X
Individual to the HygroMet and HC2 probe
Device Name
X
X
User defined description
The device name of the HC2 probe is not displayed by HW4 and is
replaced with the HygroMet Input Name
Calculation
X
X
Psychrometric calculation
HygroMet setting overrides HC2 probe setting
Data refresh rate
X
This setting has no effect on the HygroMet and probe. Depending
on the model, the data refresh rate is as follows:
HygroMet 3-wire: typically 1 s
Simulator function
X
X
Generates fixed humidity and / or temperature value
When enabled, the HygroMet settings override the HC2 probe
settings