background image

MicroMag User Manual

– March 2010

Page 15 of 23

Figure 4-1: MicroMag2-on-Carrier Reference Block Diagram

4.3 Line Descriptions

4.3.1 SCLK (SPI Serial Clock)

The serial clock input is used to synchronize both the data in and out through the

MISO and MOSI lines. The serial clock signal is provided by the customer-supplied

master device and should be 1 MHz or less. One byte of data is exchanged over

eight clock cycles. Data is captured by the master device on the rising edge of

SCLK. Data is shifted out and presented to the MicroMag2 on the MOSI pin on the

falling edge of SCLK.