beautypg.com

Dwyer instruments, inc – Dwyer CDTR User Manual

Page 4

background image

CALIBRATING SENSOR
Step 1: Slide probe into sealed calibration tube.
Step 2: Tighten screws to lock probe in place.
Step 3: Attach gas to the calibration tube.
Step 4: Follow the steps in the accessing parameter section to access the

calibration parameter (CAL).

Step 5: Press the button.
Step 6: Flow zero reference gas at 0.3 SLPM for 5 minutes.
Step 7: Press and hold the button for 3 seconds.
Step 8: Flow the full scale reference gas at 0.3 SLPM for 5 minutes.
Step 9: Press and hold the button for 3 seconds.
Step 10: Exit the parameter menu.
Step 11: Loosen screws to release probe.
Step 12: Remove probe from sealed calibration tube.

MAINTENANCE/REpAIR
Upon final installation of the Series CDT/CDTR, no routine maintenance is
required. The Series CDT/CDTR is not field serviceable and should be returned if
repair is needed. Field repair should not be attempted and may void warranty.

WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping
the product back for repair. Be sure to include a brief description of the problem
plus any additional application notes.

DWYER INSTRUMENTS, INC.

phone: 219/879-8000

www.dwyer-inst.com

p.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

fax: 219/872-9057

e-mail: [email protected]

©Copyright 2013 Dwyer Instruments, Inc.

Printed in U.S.A. 10/13

FR# 06-443800-10 Rev. 5

RESISTANCE VS TEMpERATURE TABLE

°C
-55
-50
-45
-40
-35
-30
-25
-20
-15
-10
-5
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
105
110
115
120
125
130
135
140
145
150

°f

-67.0
-58.0
-49.0
-40.0
-31.0
-22.0
-13.0

-4.0

5.0

14.0
23.0
32.0
41.0
50.0
59.0
68.0
77.0
86.0
95.0

104.0

113.0

122.0
131.0
140.0
149.0
158.0
167.0
176.0
185.0
194.0
203.0
212.0
221.0
230.0
239.0
248.0
257.0
266.0
275.0
284.0
293.0
302.0

Temperature

A

607800.00
441200.00
323600.00
239700.00
179200.00
135200.00
102900.00

78910.00
61020.00
47540.00
37310.00
29490.00
23460.00
18780.00
15130.00
12260.00
10000.00

8194.00
6752.00
5592.00
4655.00
3893.00
3271.00
2760.00
2339.00
1990.00
1700.00
1458.00
1255.00
1084.00

939.30
816.80
712.60
623.60
547.30
481.80
425.30
376.40
334.00
297.20
265.10
237.00

B

963849.00
670166.00
471985.00
336479.00
242681.00
176974.00
130421.00

97081.00
72957.00
55329.00
42327.00
32650.00
25392.00
19901.00
15712.00
12493.00
10000.00

8057.00
6531.00
5326.00
4368.00
3602.00
2986.00
2488.00
2083.00
1752.00
1480.00
1255.00
1070.00

915.50
786.60
678.60
587.60
510.60
445.30
389.60
341.90
301.00
265.80
235.30
208.90
186.10

C

289154.70
201049.80
141595.50
100943.70

72804.30
53092.20
39126.30
29124.30
21887.10
16598.70
12698.10

9795.00
7617.60
5970.30
4713.60
3747.90
3000.00
2417.10
1959.30
1597.80
1310.40
1080.60

895.80
746.40
624.90
525.60
444.00
376.50
321.00
274.65
235.98
203.58
176.28
153.18
133.59

116.88

102.57

90.30
79.74
70.59
62.67
55.83

D

78.32
80.31
82.29
84.27
86.25
88.22
90.19
92.16
94.12
96.09
98.04

100.00
101.95
103.90
105.85
107.79
109.74

111.67

113.61
115.54
117.47
119.40

121.32
123.24
125.16
127.08
128.99
130.90
132.80
134.71
136.61
138.51
140.40
142.29
144.18
146.07
147.95
149.83
151.71
153.58
155.46
157.33

E

783.2
803.1
822.9
842.7
862.5
882.2
901.9
921.6
941.2
960.9
980.4

1000.0
1019.5
1039.0
1058.5
1077.9
1097.4

1116.7

1136.1
1155.4
1174.7
1194.0

1213.2
1232.4
1251.6
1270.8
1289.9
1309.0
1328.0
1347.1
1366.1
1385.1
1404.0
1422.9
1441.8
1460.7
1479.5
1498.3
1517.1
1535.8
1554.6
1573.3

f

2394000.00
1646200.00

1145800.00

806800.00
574400.00
413400.00
300400.00
220600.00
163500.00
122280.00

92240.00
70160.00
53780.00
41560.00
32340.00
25360.00
20000.00
15892.00
12704.00
10216.00

8264.00
6722.00
5498.00
4520.00
3734.00
3100.00
2586.00
2166.00
1822.60
1540.00
1306.40

1112.60

951.00
815.80
702.20
606.40
525.60

N/A
N/A
N/A
N/A
N/A

Resistance Curves (in Ohms)

Series
Range

Configuration
CO

2

Output

Temperature
Output

Options

CDT -2

2
5

D

D

4

0
4
A
B
C
D
E
F

-LCD

RLY

4

4

Example: CDT-2D44-LCD
0 to 2000 ppm CO

2

range

0 to 5000 ppm CO

2

range

Duct
4 to 20 mA / 0 to (5 or 10) VDC
None
4 to 20 mA / 0 to (5 or 10) VDC
10 KΩ NTC thermistor type III
10 KΩ NTC thermistor type II
3 KΩ NTC thermistor
Pt100 Ω RTD
Pt1000 Ω RTD
20 KΩ NTC thermistor
Relay

figure 5: CDT Model Chart

figure 8: Resistance vs Temperature

figure 7: Elevation Chart

ft

0

400
800

1200
1600
2000
2400
2800
3200
3600
4000
4400
4800
5200
5600
6000
6400
6800
7200
7600
8000
8400
8800
9200
9600

10000

in Hg
29.92
29.50
29.10
28.69
28.29
27.90
27.51
27.13
26.76
26.39
26.02
25.66
25.30
24.95
24.60
24.26
23.93
23.60
23.27
22.94
22.63
22.31
22.00
21.70
21.40
21.40

m

0

100
200
300
400
500
600
700
800
900

1000

1100

1200
1300
1400
1500
1600
1700
1800
1900
2000
2100
2200
2300
2400
2500

hpa

1013
1002

990
979
968
957
946
935
924
914
904
893
883
873
863
853
844
834
824
815
806
797
787
779
770
761

US Customary Units

SI Units

figure 6: CDTR Model Chart

Series
Range

Configuration
CO

2

Output

Temperature
Output

RH Output
Options

CDTR -2

2
5

D

D

A

0
A
B
C
D
E
F

-LCD

RLY

4

4

Example: CDTR-2D4A4-LCD
0 to 2000 ppm CO

2

range

0 to 5000 ppm CO

2

range

Duct
4 to 20 mA / 0 to (5 or 10) VDC
None
10 KΩ NTC thermistor type III
10 KΩ NTC thermistor type II
3 KΩ NTC thermistor
Pt100 Ω RTD
Pt1000 Ω RTD
20 KΩ NTC thermistor
4 to 20 mA / 0 to (5 or 10) VDC
Relay

4

4

This manual is related to the following products: