Follow these instructions for safe and reliable operation of your GaugeCalXP Pressure Comparator.
1 Mount the comparator to your bench or table:
At higher pressures the force required to rotate the handle may cause the base of the comparator to lift.
Baseplate
16.25
(0.34)
19.4
(0.76)
Ш 10.00 (Ш 0.39), 4 places
Maximum bolt size: 8mm or 5/16"
174.9
(6.88)
142
(5.60)
390
(15.35)
428.8
(16.88)
All dimensions in mm (in)
Baseplate dimensions and suggested bolt sizes
2 Install the CPF-compatible fi ttings and any applicable adapters:
Crystal Pressure Fittings (CPF) simplify your connections and create fi nger-tight, leak-free connections up to 10 000 PSI/700 bar/70MPa working pressure.
No sealing tape is required. See the diagram below, or
, for more information on CPF.
Your GaugeCalXP is fi tted with a CPF base (7/16-20 MP) and will connect with the two included CPF male fi ttings (MPM-MPM) and two included 1/4" female
fi ttings (MPF-1/4FPT). Should you have a need for additional fi ttings,
on the following page for connection information.
Finger-tight seal uses a
Viton 90 o-ring for a seal up to
10 000 PSI / 700 bar / 70MPa
(-20 to 50oC)
Wrench-tight creates a
metal/metal seal up to
10 000 PSI / 700 bar / 70MPa
(-40 to 150oC)
point of contact