beautypg.com

Operation – Crystal GaugeCalXP User Manual

Page 4

background image

Operation

2

GaugeCalXP Operation Manual

Operation

Follow these instructions for safe and reliable operation of your GaugeCalXP Pressure Comparator.

1 Mount the comparator to your bench or table:

At higher pressures the force required to rotate the handle may cause the base of the comparator to lift.

Baseplate

16.25
(0.34)

19.4

(0.76)

Ш 10.00 (Ш 0.39), 4 places

Maximum bolt size: 8mm or 5/16"

174.9

(6.88)

142

(5.60)

390

(15.35)

428.8

(16.88)

All dimensions in mm (in)

Baseplate dimensions and suggested bolt sizes

2 Install the CPF-compatible fi ttings and any applicable adapters:

Crystal Pressure Fittings (CPF) simplify your connections and create fi nger-tight, leak-free connections up to 10 000 PSI/700 bar/70MPa working pressure.

No sealing tape is required. See the diagram below, or

visit the website

, for more information on CPF.

Your GaugeCalXP is fi tted with a CPF base (7/16-20 MP) and will connect with the two included CPF male fi ttings (MPM-MPM) and two included 1/4" female

fi ttings (MPF-1/4FPT). Should you have a need for additional fi ttings, visit

crystalengineering.net

or contact your Crystal Distributor. Refer to the diagram

on the following page for connection information.

Finger-tight seal uses a
Viton 90 o-ring for a seal up to
10 000 PSI / 700 bar / 70MPa
(-20 to 50oC)

Wrench-tight creates a
metal/metal seal up to
10 000 PSI / 700 bar / 70MPa
(-40 to 150oC)

point of contact