Operating and safety instructions, Connections – Crystal T-975-CPF Pneumatic Hand Pump User Manual
Page 7

Operating and Safety Instructions
5
T-900 Series Pump Operation Manual
Operating and Safety Instructions
CONNECTIONS
Reference Pressure Port Connections
The reference indicator threads to the upper side of the calibration hand pump. A finger-tight connection is sufficient (if utilizing an AMETEK Jofra quick
connector or a CPF fitting). If adapters are used, bonded seals and Teflon tape may be necessary.
AMETEK Jofra fitting connection.
Crystal CPF fitting connection.
Device-Under-Test Pressure Port Connections
In order to adapt the different connection threads of the device-under-test, the pressure hose can be fitted with different adapters. Please use a suitable
sealing gasket or Teflon tape as applicable for the thread type.
!
CAUTION: Do not use teflon tape with BSP or CPF threads; this may damage your hand pump.
!
CAUTION: The T-900 Series hand pump must not be soiled, or come into contact with fluids or aggressive media.
!
CAUTION: To prevent leaks, tighten the tube fitting or CPF connection to a maximum torque of 15 N-m = 11 lb-ft.