11

5. When

CAL COn” displays, press ENTER key to view the

default cell constant at 100%.
Press ENTER key again to show
the measured conductivity value.
Use ▲ and ▼ keys to scroll to
the standard’s Conductivity value
at 25

o

C.

6. Press

ENTER key to calibrate

the Conductivity value shown.
Display will show the new cell
constant being displayed. This
value will be between 70% to
130%. Otherwise an error
message “CAL“ and “Err
alternatively blinks to indicate
out of calibration range.

7. Press

ENTER key to complete

the calibration, and you will
return to “CAL” and “COn”
menu.

8. Press

MODE key and it brings

you to the Measure mode.

NOTE : Gently stir the electrode in
a container filled with calibrating
standard, and ENSURE that no
bubble is trapped during the
calibration process to avoid
erroneous reading.

IMPORTANT

: The correct

Temperature Coefficient should be
selected prior to Conductivity
calibration. (Refer to Section 5.5 for
Setting Temperature Coefficient).

SETUP

MOD

ENTE

CAL

SETUP

MOD

ENTE

COn

SETUP

MO

EN

100

SETUP

MOD

ENT

CEL

SETUP

MO

EN

CEL

SETUP

MO

EN

99