Setting up the controller, 1 setting and changing the password – Thermo Fisher Scientific Alpha CON 100 User Manual
Page 10
![background image](/manuals/370707/10/background.png)
7
4. Setting Up the Controller
4.1 Setting and Changing the Password
4.1.1 Setting
New
Password
To set a password, press MODE key
until the “SEt” is displayed.
1. Press ENTER key and the
display shows "CAL” and “Con”
alternately. Press ▲ key once
and the display shows “Set”
“id”.
2. Press ENTER to enter your
desired password. Follow the
steps below (e.g. using a
password “123”).
3. Initially, the display shows "000"
with the first digit blinking.
4.
Use ▲ (UP/INCREMENT) key to
enter “1” on the 1st digit of the
password. Press ENTER once
and the second digit will blink.
Similarly, enter the second digit
of the password “2” and press
ENTER key again. Repeat with
the third digit accordingly.
5. Press ENTER key and the
display shows “CAL” and ”COn”
alternately. Use ▲ or ▼ key to
get into the “SEt” and “id” menu,
press ENTER and set your
desired password by following
the above steps.
6. Once completed, press ENTER to confirm and then press
MODE key to return to the Conductivity display.
SETUP
MODE
ENTER
CAL
SETUP
MODE
ENTER
Con
SETUP
MODE
ENTER
SEt
SETUP
MODE
ENTER
id
SETUP
MODE
ENTER
123