beautypg.com

Thermo Fisher Scientific Alpha COND 200 (In the process of updating the brand guideline & product image) User Manual

Page 30

background image

26

10 SPECIFICATIONS

Conductivity Range

Resolution

Cell Constant

0.00 – 20.00 µS/cm

0.01 µS/cm

0.1

0.0 – 200.0 µS/cm

0.1 µS/cm

0.1

0.0 – 200.0 µS/cm

0.1 µS/cm

1

0 – 2000 µS/cm

1 µS/cm

1

0.00 – 10.00 mS/cm

0.01 mS/cm

1

0.00 – 20.00 mS/cm

0.01 mS/cm

1

0.0 – 200.0 mS/cm

0.1 mS/cm

1

0.0 – 200.0 mS/cm

0.1 mS/cm

10

Relative Accuracy

± 1 % of full scale

Temperature

-10 to 110

o

C

Resolution/Accuracy

0.1 / ± 0.5

o

C

Sensor

Pt 100 (3-wire)

Temperature Compensation

Automatic / Manual (0 to 50

°C)

Temperature Coefficient

0.0 to 5.0 %

Set-point And Controller Functions

Function Limit

Control

Switching Conductivity Hysteresis

5% of full scale

Contact Outputs, Controller

2 SPDT relays

Switching Voltage / Current / Power

Max 250 VAC / Max 3A / Max 600 VA

Signal Output / Load

4-20 mA galvanically isolated / 200 Ω

Electrical Data And Connections

Power Requirements

85 to 260 VAC, 50/60 Hz or DC

Rated Input Current (MAX)

100 mA

Insulation Category

II

Connection Terminals

3 Detachable connectors

(3-pin; 7-pin & 9-pin terminal blocks) +

1 FIXED CONNECTOR

Main Fuse

250 mA, Anti-surge (BUSSMAN S504+250mA)

Environmental Conditions

Ambient Temp. Operating Range

- 10 to 50

o

C (14 to 122

o

F)

Rel. Humidity

10 to 95 % (non-condensing)

Mechanical Specifications

Dimensions (Panel Housing – W x H x D)

96 x 48 x 150 mm

Weight

300g (350g boxed)