8

MEASUREMENT MODE

Press MODE to toggle between:

Conductivity measurement mode and

Temperature measurement mode


Conductivity Measurement Mode
The controller displays the selected Conductivity range number, R X (X ranges
from 1 to 8), for 2 seconds before displaying the Conductivity measurement.

Conductivity Range

Range No., R

Resolution

Cell Constant

0.00 – 20.00 µS/cm

1

0.01 µS/cm

0.1

0.0 – 200.0 µS/cm

2

0.1 µS/cm

0.1

0.0 – 200.0 µS/cm

3

0.1 µS/cm

1

0 – 2000 µS/cm

4

1 µS/cm

1

0.00 – 10.00 mS/cm

5

0.01 mS/cm

1

0.00 – 20.00 mS/cm

6

0.01 mS/cm

1

0.0 – 200.0 mS/cm

7

0.1 mS/cm

1

0.0 – 200.0 mS/cm

8

0.1 mS/cm

10


Temperature Measurement mode
Press MODE key once to view the temperature measurement. The display shows
ATC (Automatic Temperature Compensation) or MTC (Manual Temperature
Compensation), then the current measured temperature (for ATC) or the set
temperature (for MTC). The

o

C annunciator lights when you are measuring

temperature.

NOTE: After pressing the MODE key to display Temperature, if there is no further
key-press, the Controller will automatically revert to Conductivity Measurement
mode after about 30 seconds.

See Setup program P4.0 for further instructions.