Thermo Fisher Scientific Alpha COND 200 (In the process of updating the brand guideline & product image) User Manual
Page 12

8
MEASUREMENT MODE
Press MODE to toggle between:
•
Conductivity measurement mode and
•
Temperature measurement mode
 
Conductivity Measurement Mode 
The controller displays the selected Conductivity range number, R X (X ranges 
from 1 to 8), for 2 seconds before displaying the Conductivity measurement. 
 
Conductivity Range
Range No., R
Resolution
Cell Constant
0.00 – 20.00 µS/cm
1
0.01 µS/cm
0.1
0.0 – 200.0 µS/cm
2
0.1 µS/cm
0.1
0.0 – 200.0 µS/cm
3
0.1 µS/cm
1
0 – 2000 µS/cm
4
1 µS/cm
1
0.00 – 10.00 mS/cm
5
0.01 mS/cm
1
0.00 – 20.00 mS/cm
6
0.01 mS/cm
1
0.0 – 200.0 mS/cm
7
0.1 mS/cm
1
0.0 – 200.0 mS/cm
8
0.1 mS/cm
10
 
Temperature Measurement mode 
Press MODE key once to view the temperature measurement. The display shows 
ATC (Automatic Temperature Compensation) or MTC (Manual Temperature 
Compensation), then the current measured temperature (for ATC) or the set 
temperature (for MTC). The 
o
C annunciator lights when you are measuring
temperature. 
 
NOTE: After pressing the MODE key to display Temperature, if there is no further 
key-press, the Controller will automatically revert to Conductivity Measurement 
mode after about 30 seconds. 
 
See Setup program P4.0 for further instructions. 