P3.0 rnge p3.1, K 10 – Thermo Fisher Scientific Alpha COND 200 (In the process of updating the brand guideline & product image) User Manual
Page 24
![background image](/manuals/370695/24/background.png)
20
7.5 Measurement Range Selection – P3.0
Setup program P3.0 is for selecting the range of measurement.
P3.1: select measurement range and corresponding cell constant
P3.0
RNGE
P3.1
2000
K 10
.
Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P3.1: Select Measurement Range
Set controller to the specific range of measurement. Ensure the cell you have
connected to the controller has the same cell constant as that stated in the range.
1.
Key in the password “022” as per procedure in Section 5.
2.
Press ▲key until screen displays P3.0 and RNGE. Press ENTER.
3.
The screen will scroll P3.1 and the selected range.
4.
Press ▲or ▼keys to select appropriate range and cell.
Available ranges are as follows (Default range is highlighted):
Conductivity Range
Cell Constant
4 mA current 20
mA
current
0.00 – 20.00
µS/cm
0.1 0.00
µS/cm
20.00
µS/cm
0.0 – 200.0
µS/cm
0.1 0.0
µS/cm
200.0
µS/cm
0.0 – 200.0
µS/cm
1 0.0
µS/cm
200.0
µS/cm
0 – 2000 µS/cm
1 0
µS/cm
2000
µS/cm
0.00 – 10.00
mS/cm
1 0.00
mS/cm
10.0
mS/cm
0.00 – 20.00
mS/cm
1 0.00
mS/cm
20.0
mS/cm
0.0 – 200.0
mS/cm
1 0.0
mS/cm
200.0
mS/cm
0.0 – 200.0
mS/cm
10 0.0
mS/cm
200.0
mS/cm
5.
Press ENTER to confirm (The set points 1 & 2, Conductivity and
Temperature calibration are reset to default values).
6.
Press ▲and ▼keys together (ESCAPE) to return to P3.0. Press the ▲and
▼keys together again, to return to measurement mode.