7.6 Configure Temperature Settings – P4.0
Setup program P4.0 is for selecting ATC or MTC, set temperature coefficient
values, and select normalization temperature
P4.1: ATC or MTC mode (default : ATC mode)
P4.2: set temperature coefficient (default : 2.10 %)
P4.3: set Normalization temperature (default : 25.0

o

C)

P4.0

P4.1

P4.2

P4.3

2

5

0

TEMP

COEF

NORM

MODE

MODE

T

A C

T

M C

0

2

.

.


P4.1: Selecting ATC or MTC
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P4.0 and TEMP. Press ENTER.

3.

Screen will scroll P4.1, MODE, then either ATC or MTC.

4.

Press the ▲or ▼keys to select either ATC or MTC. Press ENTER.

5.

Proceed to P4.2 or press the ▲and ▼keys together to return to P4.0. Press
the ▲and ▼keys together again, to return to Conductivity measurement
mode.


P4.2: Setting Temperature Coefficient
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P4.0 and TEMP. Press ENTER until
screen shows P4.2.

3.

Screen will scroll P4.2, COEF., then 2.1 (to indicate 2.1%).

4.

Press the ▲or ▼keys to change the temperature coefficient value. Press
ENTER to confirm.

5.

Proceed to P4.3 or press the ▲and ▼keys together to return to P4.0. Press
the ▲and ▼keys together again, to return to Conductivity measurement
mode.


P4.3: Setting Normalization Temperature
1.

Key in the password “022” as per procedure in Section 5.

2.

Press ▲key until screen displays P4.0 and TEMP. Press ENTER until
screen shows P4.3.

3.

Screen will scroll P4.3, then NORM, then 25.0.

4.

Press ▲or ▼keys to change the normalization temperature value. Press
ENTER to confirm.

5.

Press ▲and ▼keys together (ESCAPE) twice, to return to measurement
mode.