26

7.4.2

Selecting pH or ORP mV measurement


Besides selecting the controller to function as a pH or ORP (mV) controller, it

also allows the selection of the type of input: symmetrical or asymmetrical.

Symmetrical mode is used in conjunction with a potential matching pin – in an

electrically noisy environment. Please refer to section 3, page 6 for a graphical

representation.

1) Follow directions in 7.4.1 to enter Configuration mode.

2) Press ENTER key. Scroll with

Δ or keys until upper display shows mode of

operation (pH or ORP mV) and lower display shows symmetrical (SY) or

asymmetrical (ASY) input type.

3) Proceed to 7.4.3, or return to Measurement mode by pressing

Δ and keys

simultaneously (escape).


7.4.3

Reverting to factory default settings

Use this parameter to reset all settings to factory default. Changing from “no” to

“YES” and pressing ENTER key resets all settings to factory default.

WARNING: If “Yes” is selected, all settings

input will be overwritten!

1) Follow directions in 7.4.1 to enter Configuration

mode.

2) Press ENTER key. Scroll with

Δ or keys until upper display shows “no” or”

YES”, and lower display shows “deF” (default).

3) Press

Δ or keys to select no or yes. Selecting “YES” and pressing

ENTER key will overwrite all user-settings and automatically return to

Measurement mode.

4) Press ENTER key to confirm your selection and return to Advanced Set-up

mode, or return to Measurement mode by pressing

Δ and keys simultaneously

(escape).

no

dEF

SETUP

HOLD