Thermo Fisher Scientific Alpha pH 800 User Manual

Page 21

background image

14

3) Press ENTER key to start calibration at pH 7 (or pH 6.86). The electrode

indicator and CAL indicator both flash. The controller automatically adjusts

reading to match the buffer value.

4) The lower display will now show its next lower ‘pH’ buffer. Use

Δ and keys

to select second buffer from one of 5 remaining preset values. Remove

electrode from first buffer, wash and then immerse into second buffer. Once

upper display stabilises press ENTER key, electrode and CAL indicator will

flash. The controller adjusts reading to the buffer.

5) After calibrating to a second buffer value, controller automatically displays

electrode slope and offset. The zero point and slope are re-determined after

each calibration.

6) Press ENTER key. If calibration mode was accessed using CAL key,

controller will return to Measurement mode. If calibration mode was accessed

from Advanced Set-up mode, controller will return to sub-function menu.

Note: If there is a calibration error, controller displays “ERR”. If this happens,

push both

Δ and keys

(escape) to restart calibration beginning from step 1.

Note: When calibrating with manual temperature compensation, controller

automatically changes from preset process temperature to calibration

temperature. After leaving Calibration mode, controller switches back to process

temperature (for setting calibration temperature and process temperature, see

section 7.2.3).