13

6 CALIBRATION

Direct Calibration from Measurement mode is possible via the CAL key.

Calibration procedure can also be accessed from Advanced Setup mode.

6.1 pH

Calibration

This unit features five preset buffer values (1.00, 4.00, 6.86 or 7.00, 9.00, 9.18

and 10.00) for quick, two-point auto calibration. When calibrating this instrument,

standard pH buffer solutions must be used.

1)

Enter Calibration mode. While in

Measurement mode, push CAL key. Upper and lower

display reads “CAL” & “pH”. Push ENTER or CAL

key again. The controller displays electrode status

based on previous calibration. Electrode slope and

offset information is displayed.

NOTE: If display reads “CAL OrP”, see section 7.4

for procedures on how to switch from ORP to pH

readings.

2)

Press ENTER key again to begin calibration.

The “CAL” indicator appears at top of display and

buffer indicator appears in bottom left corner of

display. Upper display is on-line pH value as

measured by electrode. Lower display indicates the

first compulsory calibration value, pH 7.00 (or

6.86).

Immerse clean electrode in buffer solution. In

ATC, temperature probe must also be immersed

in buffer solution. Likewise for symmetrical

mode, immerse potential matching pin in buffer.