Temperature calibration (ATC mode only)


1) Select “ATC on” as described above in

Section 7.2.1.

2) Press ENTER key. The upper display

indicates current temperature offset. The

current measured temperature is shown in the

lower display.

3) Compare current measured temperature

on the controller display to a thermometer

known to be accurate. Note down the correct

temperature value.

4) Press

Δ or keys to scroll the lower display to match the correct value.

Upper display will now show the offset value. You can offset temperature up to

±

5

o

C.

5) Press ENTER key to confirm your selection.

6) Continue with additional Advanced Set-up procedures, or return to the

Measurement mode by pressing

Δ and keys (escape) simultaneously.