ADVANCED SET UP MODE

7.1

Electrode Offset (OFS) sub-function

This option is available only in the pH mode. An allowance of

±120mV is allowed

to correct for electrode deviations. This feature is useful for prolonged on-line

applications.

The controller will add or subtract the value from the measured pH and display

the correct value. Should the offset value be too high, consider performing

maintenance on the electrode followed by calibration. If the calibration is

unsuccessful, please replace the electrode.

1) Take a sample from the system and note pH reading

displayed on controller.

2) Measure pH of sample using a calibrated pH tester,

hand-held meter, or bench meter. Record the correct pH

value.

3) Enter Advanced set-up mode. Press ENTER.

4) Press

Δ or keys to scroll until upper display shows

OFS.

5) Press ENTER key. The SETUP annunciator appears at the top of display.

Lower display shows on-line measured pH value. Upper display shows current

offset value.

6) Press

Δ or keys until pH value coincides with correct pH value noted in

Step 2. As the

Δ or keys is pressed to adjust pH value, the ‘offset’ value in

upper display also changes.

7) Press ENTER key to enter offset value.

SETUP

SETUP

HOLD

HOLD

Ofs

-0.05

7.00