Instruction Manual

Silver/Silver Sulfide Electrode

25

Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


ISA not used

use recommended ISA


standard used as ISA

use ISA


electrode exposed

check section entitled

to complexing agents

Complexation


air bubble on membrane

remove bubble by

re-dipping probe



"Incorrect Answer"

incorrect scaling of

plot millivolts on the linear

(but calibration

semi-log paper

axis. On curve is good)

the log axis, be sure that

concentration

numbers

within

each

decade

are

increasing

with

increasing

concentration


incorrect sign

be sure to note sign of

millivolt

number

correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion

factor:

1.0X10

-3

M= 32.1 ppm S

-2

= 2.0X10

-3

N (S

-2

);

= 108 ppm Ag

+1


complexing agents

check section entitled

in

sample

Complexation