Silver/Silver Sulfide Electrode

Instruction Manual

24

TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with

Reading shorting

strap

(see

meter

instruction

manual)


defective electrode

check electrode operation

electrodes

not unplug

electrodes

plugged in properly

and reset


reference electrode

be sure reference

not filled

electrode is filled


air bubble on

remove bubble by

membrane

re-dipping

electrode


electrodes

not put

electrodes

in

solution

in

solution



Noisy or Unstable

defective meter

check meter with

Readings

(readings

shorting

strap

continuously or

rapidly changing)

air bubble on

remove bubble by

membrane

re-dipping

electrode


ISA not used

use recommended ISA


meter or stirrer

ground meter or stirrer

not grounded


defective

electrode

replace

electrode


electrode exposed

soak electrode in

to interference

silver or sulfide standards


Drift (reading

samples and standards

allow solutions to

slowly changing

at different temperatures

come to room temperature

in

one

direction) before

measurement


electrode exposed

check section entitled

to complexing agents

Complexation


incorrect

reference

use

recommended

filling

solution filling

solution