Calcium Electrode

Instruction Manual

18

between samples



Drift (reading

samples and standards

allow solutions to come to room

slowly changing

at different

temperature before measurement

in one direction)

temperatures


electrode exposed

soak electrode in

to interferences

calcium standard


incorrect reference

use recommended

filling solution

filling solution


incorrect pH

adjust to pH 3-10

with NaOH or HCl


Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


defective electrode

check electrode operation



air bubble on membrane

remove bubble by

re-dipping probe


electrode exposed to

soak electrode in

interferences

calcium standard


standard used as ISA

use ISA


ISA not used

use ISA