Thermo Fisher Scientific Ion Selective Electrodes Calcium User Manual

Page 17

background image

Instruction Manual

Calcium Electrode

17

TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with shorting strap

Reading

(see meter instruction manual)


defective electrode

check electrode operation


electrodes not

unplug electrodes and reset

plugged in properly


reference electrode

be sure reference electrode is filled

not filled


air bubbles on

remove bubbles by

membrane

re-dipping electrode


electrodes not

put electrodes in solution

in solution


Noisy or Unstable

defective meter

check meter with

Readings (readings

shorting strap

continuously or
randomly changing air bubble on

remove bubble by

membrane

re-dipping electrode


meter or stirrer

ground meter or stirrer

not grounded


outer filling solution

fill electrode to level just

too low

below fill hole


defective electrode

replace electrode


electrode exposed to

soak electrode in calcium standard

interferences


"Incorrect

incorrect scaling

plot millivolts on the linear

Answer"

of semi-log paper

axis. On the log axis, be sure

(but calibration

concentration numbers within

curve is good)

each decade are increasing

increasing concentration


incorrect sign

be sure to note sign of

millivolt number correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

1.0x10

-3

M = 40 ppm Ca

+2

=

100 ppm as CaCO

3

sample carryover

rinse electrodes thoroughly