Thermo Fisher Scientific Eutech DO 700 User Manual
Page 28

24
PROBLEM CAUSE
SOLUTION
CAL
% calibration value is not
acceptable.
Electrode is disconnected
or failing.
Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.
CAL
ppm or mg/L calibration is
attempted when the factory
calibrated absolute value is
below 2.00
Use concentration calibration
solution higher than 2.00.
Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.
Slow
response or
unstable
readings
External noise pickup or
induction caused by
nearby electric motor
Insufficient electrolyte in
probe
Air bubbles near sensor
Dirty or damaged probe
Probe not deep enough in
sample
Broken probe
Remove or switch off interfering
device
Fill probe with electrolyte & / or
replace pre-assembled cap
Stir or tap probe to remove
bubbles, reduce stirring
Clean the probe and re-calibrate
Submerse sensor cap into sample
Replace probe or membrane
- PCTestr 35 (2 pages)
- pHScan BNC (3 pages)
- pHScan 3/3+ (5 pages)
- pHTestr 1 (3 pages)
- pHTestr 10/20/30/10 BNC/Spear (2 pages)
- ORPTestr 10/10 BNC (2 pages)
- EC/TDS/SaltTestr 11 (4 pages)
- EC/TDS/SaltTestr (2 pages)
- ECScan High/Low & TDScan High/Low (9 pages)
- SaltTestr (2 pages)
- EcoTestr pH 2 (2 pages)
- EcoTestr pH 1 (2 pages)
- EcoTestr EC High (2 pages)
- EcoTestr EC Low (2 pages)
- EcoTestr TDS High (2 pages)
- EcoTestr TDS Low (2 pages)
- EcoTestr Salt (2 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (New version R1.1, SN >797406) (23 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (Old version EP6, SN <797406, discontinued) (23 pages)
- Eutech COND/TDS/Salt 6 Plus (40 pages)
- Eutech DO 6 Plus (48 pages)
- EcoScan pH/Ion 5 & 6 (27 pages)
- EcoScan CON 6 & TDS 6 (56 pages)
- EcoScan CON 5 & TDS 5 (18 pages)
- EcoScan Salt 6 (40 pages)
- EcoScan DO 6 (80 pages)
- CyberScan pH 10/pH 100 (67 pages)
- CyberScan pH 11/pH 110 (76 pages)
- CyberScan CON 10/CON 100/CON 200 (62 pages)
- CyberScan CON 11/CON 110 (80 pages)
- CyberScan DO 110 (60 pages)
- CyberScan PCD 650 (127 pages)
- CyberScan CON 400/410 (For units manufactured before March 2010, discontinued) (60 pages)
- CyberScan CON 400 (For units manufactured from March 2010 onwards) (60 pages)
- CyberScan pH 300/310 (52 pages)
- CyberScan DO 300 (60 pages)
- CyberScan PC 300 (72 pages)
- CyberScan PD 300 (76 pages)
- CyberScan PC 10 (31 pages)
- C401 Colorimeter (64 pages)
- TN100 Turbidimeter (31 pages)
- RS232C Interface Adapter (9 pages)
- Thermo Scientific Temp 360 (44 pages)
- Thermo Scientific Temp 340 (40 pages)
- Thermo Scientific Temp 300 (32 pages)