6) Press

to confirm the calibration. The primary display will blink

briefly before returning to measurement mode.

NOTE: The minimum calibration value is 2 mg/L. In addition, the
calibration is limited to ±40% adjustment of the factory default value to
prevent erroneous calibrations.

5. SETUP

FUNCTIONS

Use the setup feature to customize your instrument operation.

There are two setup menus; one menu is derived from the %
saturation mode and the other from the mg/L (or ppm) concentration
mode.

First, select the appropriate measurement mode you wish to adjust by

pressing

until the desired units are displayed. During

measurement, press and hold

for 5 seconds to enter SETUP

mode of the parameter being measured.

Press

or

to change programs or change options.

Press

to select the program or confirm selection.

Press

to go back one level or return to measurement mode.

5.1 1.0 CAL (Calibration)

Press

to view stored calibration points.

Note: % saturation mode will list two points, while
concentration mode has only one point.