5. PROBE CARE AND MAINTENANCE

Keep your Conductivity probe clean. Do not strike the probe against any hard
surface, and never scratch the stainless steel pins with hard material.

Wash the probe thoroughly with tap or deionised water after each use.

Do not immerse the probe in oily solutions. To remove oil films or oxidation,
clean electrode with alcohol.

Clean it with a mild detergent solution. Blot it dry. Wash thoroughly in tap water
and then in deionised water. Recalibrate the meter after cleaning it.

6. TROUBLESHOOTING

Problem

Cause

Solution

No display

a) Batteries not in

place.

a) Insert batteries.

b) Re-insert batteries in correct polarity.

Unstable reading

a) Dirty electrode.

b) Broken electrode.

a) Clean electrode and recalibrate.

b) Replace electrode.

“E2” display

a) Error in calibration.

a) Recalibrate the instrument.

“E3” display

a) Failed

initialization.

a) Turn off meter and turn it on. Return if

necessary.

Not able to
calibrate

b) Display freezes.

c) Faulty electrode.

b) Release reading by pressing HOLD

key.

c) Replace electrode.