Ca 1395 – Thermo Fisher Scientific EcoScan CON 5 & TDS 5 User Manual

Page 10

background image

EcoScan CON 5 / TDS 5

9

4. Press

∆∆∆∆

(up) or

(down) key to scroll the reading until the displayed value

matches your standard solution i.e. 1413

µ

S/cm.

5. Press

ENTER key to confirm calibration and the LCD displays the new set

reading and revert to measurement mode.

6. Alternatively you can abort the new calibrated value by pressing CAL key to

exit from calibration mode.

Important: Meter allows a tolerance range of 30% of its measured reading. You
will not able to scroll through to the desired value if reading is out of its
tolerance.

3.2 TDS Calibration (EcoScan TDS 5)

The calibration procedure for the EcoScan TDS 5 is the same as EcoScan CON
5 (refer to section 3.1 for details).

CA

1395

uS

1413

uS