beautypg.com

Thermo Fisher Scientific EcoScan CON 5 & TDS 5 User Manual

Page 12

background image

EcoScan CON 5 / TDS 5

11

3.3.1.1 With Temperature probe (with Curve Selection)

In the event the temperature probe has drifted too far from its original
characteristic – due to age and use or if the probe is being replaced, it may be a
good idea to match the probe to the best curve. There are 3 curves
programmed in the unit. To choose the curve, proceed as follows:

1. Connect your temperature probe to the meter. Press MODE key to enter

Temperature mode until “

o

C” annunciator appears in the LCD.

2. Compare the displayed value to a NIST certified thermometer or other

thermometer known to be accurate. For best accuracy, place probe and
thermometer in a constant temperature bath.

3. Press CAL key to enter temperature calibration mode. The LCD shows

“CA” momentarily and the displayed reading flashes.

4. Press

MODE key once. Note the reading. This is with respect to the curve

1.

5. Press

∆∆∆∆

(up) key once. The display will change to a different value which is

with respect to curve 2. Note the reading. Press

∆∆∆∆

(up) key again, the

reading now will be with respect to curve 3. Choose the curve which gives
its reading closest to the actual value.

6. Press

ENTER key and the display blinks once. The curve selection has now

been made. To match the value exactly with the standard value, press

∆∆∆∆

(up) or

(down) key, the display can be set to the exact value. Press

ENTER to confirm.