Techne DB-2TC User Manual
Page 22

THE DB-2TC DRI-BLOCK
®
The unit has a recessed chamber which can hold up to two interchangeable insert blocks which accept
test tubes or other sample containers. The temperature is set by 3 front panel buttons with digital display.
The DB-TC is supplied with an extraction tool for removing the insert blocks from the unit.
Working temperature range
Ambient 0°C to 100°C
Temperature range
0°C to 100°C
Temperature stability @ 37°C
±0.1°C
Temperature setting
Push button
Temperature scale graduation
No graduation
Temperature display
Red LED 4 digit
Uniformity within block at 37°C
±0.1°C
Uniformity within block at 100°C
±0.1°C
Block at maximum temperature
1°C
Display resolution
0.1°C
Set point accuracy
±1°C
Maximum number of blocks
2
Heater power
2 x 150W
Heat up time 30°C to 37°C
6 minutes
Heat up time 30°C to 56°C
14 minutes
Heat up time 30°C to 100°C
19 minutes
Overall size L x W x H (mm)
279 x 260 x 105
Net weight (without blocks)
2.9 kg
Electrical supply
Voltage
Frequency
Power
230V
50Hz - 60Hz
300W
110V - 120V
50Hz - 60Hz
300W
20
®