THE DB-3D DRI-BLOCK

®

This unit has a recessed chamber which can hold up to three inter-changeable insert blocks which accept
test tubes or other sample containers. The temperature is set by 3 front panel buttons with digital display.

The DB-3D is supplied with an extraction tool for removing the blocks from the unit.

Working temperature range

Ambient 0°C to 200°C

Temperature range

0.0°C to 200.0°C

Temperature stability

±0.1°C at 40°C

Temperature stability

±0.15°C at 100°C

Temperature setting

3 push buttons

Temperature display

4 digit LED

Set point resolution

0.1°C

Set point to accuracy

±1°C

Max temperature variation between blocks

0.2°C at 40°C

Heat up time 30°C to 37°C

11 minutes

Heat up time 30°C to 56°C

15 minutes

Heat up time 30°C to 200°C

25 minutes

Number of blocks

3

Overall size L x W x H (mm)

279 x 260 x 105

Electrical supply

Voltage

Frequency

Power

230V

50Hz-60Hz

650W

110V-120V

50Hz-60Hz

650W