THE DB-3D DRI-BLOCK
®
This unit has a recessed chamber which can hold up to three inter-changeable insert blocks which accept
test tubes or other sample containers. The temperature is set by 3 front panel buttons with digital display.
The DB-3D is supplied with an extraction tool for removing the blocks from the unit.
Working temperature range
Ambient 0°C to 200°C
Temperature range
0.0°C to 200.0°C
Temperature stability
±0.1°C at 40°C
Temperature stability
±0.15°C at 100°C
Temperature setting
3 push buttons
Temperature display
4 digit LED
Set point resolution
0.1°C
Set point to accuracy
±1°C
Max temperature variation between blocks
0.2°C at 40°C
Heat up time 30°C to 37°C
11 minutes
Heat up time 30°C to 56°C
15 minutes
Heat up time 30°C to 200°C
25 minutes
Number of blocks
3
Overall size L x W x H (mm)
279 x 260 x 105
Electrical supply
Voltage
Frequency
Power
230V
50Hz-60Hz
650W
110V-120V
50Hz-60Hz
650W
19