Specification – Techne DB-2TC User Manual
Page 20
SPECIFICATION
THE DB-2D DRI-BLOCK
®
This unit has a recessed chamber which can hold one or two inter-changeable insert blocks which accept
test tubes or other sample containers. The temperature is set by 3 front panel buttons under the digital
display.
The DB-2D is supplied with an extraction tool for removing the blocks from the unit and two plastic half
blocks for reducing heat losses when only one aluminium block is used.
Specification – DB-2D
Working temperature range
Ambient 0°C to 100°C
Temperature range
0.0°C to 105.0°C
Temperature stability
±0.05°C at 40°C
Temperature stability
±0.1°C at 100°C
Temperature setting
3 push buttons
Temperature display
4 digit LED
Set point resolution
0.1°C
Set point accuracy
±1°C
Max temperature variation between blocks
0.2°C at 40°C
Heat up time 30°C to 37°C
8 minutes
Heat up time 30°C to 56°C
9 minutes
Heat up time 30°C to 100°C
12 minutes
Number of blocks
2
Overall size L x W x H (mm)
202 x 260 x 105
Electrical supply
Voltage
Frequency
Power
230V
50Hz-60Hz
400W
120V
50Hz-60Hz
400W
18