Fitting Installation. EX80 Series meters require spe-

cial fittings. The meter fitting must first be installed in the

pipeline. Straight pipe of at least ten times the diameter

upstream of the meter and five diameters downstream is

strongly recommended in order to achieve proper accu-

racy. These are

minimum values. As the diagrams on the

the next page will show, you may need much more straight

run under specific adverse circumstances.

If you can't provide enough run to smooth out the turbu-

lence caused by valves, fittings, and changes in direction,

some decrease in accuracy may result. This does not

mean that the flow meter's reading is meaningless, how-

ever. In some applications, a repeatable reading may be

more important than a highly accurate one. In applica-

tions where the flow meter is part of a control system,

operating a valve or controlling chemical addition, repeat-

ability of reading is more critical than absolute accuracy.

EX80 Series PVC meter tees are supplied with some up-

stream straight pipe. The length provided may be less

than ten diameters upstream and five downstream. It is

not advisable to connect directly to the end of these

fittings with a flow-disturbing device such as a valve or

elbow. If possible, straight pipe should be added to the

upstream end of these fittings.

INSTALLATION

Caution: Never remove the u-clip
retainer when the pipe is under pressure.
Always remove pressure from the pipe

before you attempt to remove the meter. Removal
under pressure may result in damage or serious
injur y.

A PVC fitting is usually installed by solvent welding. The

stainless steel and brass meter fittings have female pipe

threads, requiring the appropriate male threaded fittings.

Saddle fittings (size 3" and above) require a hole to be cut

in the pipe. The recommended hole size is 1-3/4".

Meter Installation. After the meter fitting is installed in

the pipeline, the meter can be installed in the fitting. After

noting the direction of the flow arrow, press the meter into

the fitting as far as it will go. Retain the meter in place by

inser ting the u-pin. The pin can be installed from either

side. It may be necessar y to rotate the probe back and

for th slightly to star t the pin into the slots on the probe.

Slide the pin in as far as it will go.

5X

Diameter Minimum

(See Below)

Page 2

DISTORTED FLOWS

Distorted
Flow Profile

Faster Flow

Causes Meter

To Read High

FLOW

10X

Diameter Minimum

(See Below)

Okay position if there is

no air in the pipe

BEST POSITIONS