Ar ch ive d – Seametrics EX80 Series v.1 User Manual
Page 11
Problem
Probable Cause
Try...
No pulse output
Output pulses
incorrect
Check Plumbing
Connect to earth ground
Check for proper electrical wiring
Check for power across power
input terminals
Reverse connections
Change output connections
Select another flow meter
Check for proper ground
Check for proper electrical wiring
Select another flow meter
Check for full pipe or install meter in the
ver tical position
Check for ten diameters upstream
AND
five diameters downstream
Pipe not full
Unit not grounded
Excessive electrical noise
No power
Power Reversed
Output connections reversed
Fluid conductivity <20 microseimans/cm
Missing or incorrect ground wire
Excessive electrical noise
Fluid conductivity <20 microseimans/cm
Empty pipe
Not enough straight pipe
TROUBLESHOOTING
Page 9
AR
CH
IVE
D
(Includes Dates 6/14/04 to 5/05)
- EX800-Series (12 pages)
- EX100-200-Series (16 pages)
- IP800-Series (12 pages)
- IP100-200-SERIES (12 pages)
- TX800-Series (12 pages)
- TX100-200-SERIES (12 pages)
- PE102 (4 pages)
- SEB (2 pages)
- SES (4 pages)
- S-Series (4 pages)
- MJ-Series (4 pages)
- MJN-Series (4 pages)
- WJ-Series (4 pages)
- WT-Series Turbine (4 pages)
- iMAG-Series (16 pages)
- WMX-Series (12 pages)
- WMX-Series (8 pages)
- WMP-Series (8 pages)
- AG2000-Series (4 pages)
- AO55 (4 pages)
- FT400-Series (12 pages)
- DL76 (4 pages)
- FT500-Series (8 pages)
- FlowInspector (16 pages)
- RSP5 (4 pages)
- PT35 (4 pages)
- PT34 (4 pages)
- PS40 (2 pages)
- PD10 (4 pages)
- PC-SERIES (4 pages)
- AD30 (2 pages)
- AG1000 (10 pages)
- AG2000 v.1 (4 pages)
- AG2000 v.2 (4 pages)
- AO45 (4 pages)
- DL75 v.1 (4 pages)
- DL75 v.2 (24 pages)
- EM 100-102 v.1 (4 pages)
- EM 100-102 v.2 (4 pages)
- EM 101 v.1 (6 pages)
- EM 101 v.2 (8 pages)
- EX100-200 Series v.1 (14 pages)
- EX100-200 Series v.2 (16 pages)
- EX80 Series v.2 (12 pages)