beautypg.com

Ar ch ive d – Seametrics EX80 Series v.1 User Manual

Page 10

background image

OPERATION & MAINTENANCE

When the EX81 or 82 is powered up and there is no flow,

there should be no output pulses (or, if connected to the

FT420, flow rate should read “0”). If there are pulses...

Zero Adjustment. In some cases it may be necessary

to adjust the flow meter under no-flow conditions after it

has been installed.

To per form the adjustment, after determining that there is

no flow, short between the two pins marked “Zero Adj.” A

red LED light will come on for approximately 50 seconds

and then go out. The zero adjustment is completed.

Minimum Flow.

As with any other flow sensor, there is a

rate below which the EX80-series sensor cannot read.

Check the flow rate table below for the minimum flow rate

detectable by the sensor for a given pipe size.

Filtering.

The software of the EX80-series sensor filters

out electrical noise and also averages sudden variations

in the flow to smooth the output. It takes a matter of

seconds for the flow sensor to get up to full output when it

is powered up or when flow begins. If the unit responds

ver y slowly, it is probably due to excessive electrical noise.

If this is the case, check for adequate grounding to improve

the response.

Electrode Coating.

Grease or other adhering, non-

conductive materials can stop flow detection if the

electrodes become heavily coated. To clean the electrodes,

remove the sensor from the pipe and gently scrub the

electrodes (three dark grey bumps) on the reading face of

the flow sensor. A mild soap (dishwashing liquid for

example) can be used to aid the cleaning process.

Page 8

K-factor. If the EX80 Series meter is ordered with its

fitting, it is factor y calibrated in the fitting. A K-factor (meter

factor) is indicated on the side of the fitting. This repre-

sents the actual number of pulses per gallon the meter

produced during the factor y flow test. This number can

entered into an FT420 or FT520 flow indicator to make it

read properly. If a pulse divider is being used, the K-factor

is the star ting point for calculating the divider number.

LOCATION OF K-FACTOR

10031295

MF81T-P200

K: 53.6

Find Your K-Factor Here

1” 1-1/2" 2" 3" 4" 6" 8"

• Min .54 1.3 2 4.5 8 18 31

Max 54 127 209 461 794 1800 3120

FLOW RANGE (GPM)

AR

CH

IVE

D

(Includes Dates 6/14/04 to 5/05)