Seametrics IP800-Series User Manual
Page 8
![background image](/manuals/370225/8/background.png)
Page 8
Minimum Flow.
As with any other flow sensor, there is a rate
below which the IP800-Series sensor cannot read. Check the
flow rate table below for the minimum flow rate detectable by
the sensor for a given pipe size.
Calibration (“K-factor”).
The K-factor represents the number
of pulses per gallon the meter produces during a flow test.
This number must be entered into your electronic control to
make it read properly. If the IP800-Series meter is ordered
with a tee fitting, it is factory-calibrated in the fitting and the
K-factor is indicated on the side (see below).
10031295
MF81T-P200
K: 53.6
Find Your K-Factor Here
Field Calibration.
It is possible to field-calibrate an IP800-
Series flow sensor to determine an accurate K-factor in the
actual installation. The reason for doing this would be to
compensate for an unusual condition, for instance, applica-
tions with higher viscosity fluid (IP meters are calibrated for
water use) or which lack adequate straight pipe ahead of the
meter. Field Calibration procedures are described in a Tech-
nical Bulletin on our website (www.seametrics.com).
If a saddle or weld-type fitting has been ordered, use the K-
factor calculator at the bottom of the www.seametrics.com
home page to determine the K-factor. In PVC, however, it is
possible to order a saddle pre-installed on a standard length
of pipe, and the fitting can be wet-calibrated in this case.
OPERATION
50
100
150
200
60˚
70˚
80˚
90˚ 100˚
110˚
120˚ 130˚
140˚
P.S.I.
PRESSURE VS. TEMPERATURE (PVC/Polypro)
½
"
¾
"
1"
1½"
2"
3"
4"
6"
8"
Min
0.28
0.5
0.8
1.9
3.1
6.9
12
27
46.8
Max
28
50
80
190
314
691
1190
2700
4680
Flow Range (Gallons per Minute)
½
"
¾
"
1"
1½"
2"
3"
4"
6"
8"
Min
1.06
1.89
3.03
7.19
11.73
26.11
45.42
102.21
117.16
Max
105.99 189.27 302.83 719.22 1188.61 2615.72 4504.64 10220.61 17715.73
Flow Range (Liters per Minute)