beautypg.com

Line voltage wiring – Agilent Technologies 6613C User Manual

Page 42

background image

3 - Troubleshooting

42

Line Voltage Wiring

Figure 3-3 illustrates the primary wiring configuration of the power transformer for various ac line voltages.
Use long nose pliers to disconnect the wires going to the transformer terminals.

NOTE:

Install the correct fuse when changing the ac line voltage from a previous setting:
for 110/120 Vac: 2.5AT, 250V, Agilent p/n 2110-0633;
for 220/230 Vac: 1.25AT, 250V, Agilent p/n 2110-0788

Figure 3-3. Transformer Wiring

white/red/grey

grey

orange

orange

1
2
3
4
5
6

7

Top part of
transformer

Front of unit

white/red/grey

grey

orange

white/red/grey

grey

orange

orange

white/red/grey

grey

orange

220 VAC

(spare)

orange

(spare)

orange

white/red

white/black

white/brown

white/brown

white/black

white/red

red

black

white/yellow

white/violet

white/yellow

white/violet

white/yellow

white/violet

white/yellow

white/violet

Top part of
transformer

Front of unit

120 VAC

Top part of
transformer

Front of unit

100 VAC

Top part of
transformer

Front of unit

230 VAC

1
2
3
4
5
6

7

1
2
3
4
5
6

7

1
2
3
4
5
6

7

Bottom part of

transformer

Front of unit

All Voltages