Troubleshooting - 3 29 – Agilent Technologies 6613C User Manual

Page 29

background image

Troubleshooting - 3

29

Figure 3-1 Sheet 1. Troubleshooting Flowchart

Turn on unit and observe the

display. All of the segments and

annunciators, the address and

then after self test should display

an error message or go to the

metering mode.

Display comes on?

+5V @ A2J211-1 (to

chassis)?

Replace A3

Check A1F305,

Red/W hite/Black

cable A1-A2 &

cable A2-A3, track
on A2 (J206-J211)

Error Message?

Yes

Go to Error Message

Table 3-2.

Yes

Protect

annunciator

on?

No

RI?

Yes

OV?

No

Check for OV setting <

Voltage setting, Replace

A1

Yes

OC?

No

Check that OCP is not

enabled, Replace A1

Yes

FS?

No

Check F309 (fuse near

main heat sink),

Replace A1

Yes

For OT check fan,

Replace A1

No

Go to Sheet 2

No

No

A3J111-5 low (no

pulses)?

Yes

No

Replace A2

Yes

Check Bias voltages

(see Table 3-3)

Bias voltages OK?

Transformer Inputs

OK?

No

Yes

Check Main Fuse,

Replace T1

No

Replace A1

Yes

No