Troubleshooting - 3 31, Figure 3-1 sheet 3. troubleshooting flowchart – Agilent Technologies 6613C User Manual

Page 31

background image

Troubleshooting - 3

31

Figure 3-1 Sheet 3. Troubleshooting Flowchart

From Sheet 2

Program current to full

scale, voltage to Vmax

and load to the power

supply's rated current.

Supply should be in CC.

Will not go into CC

or error > 10%

?

Output out of spec

but close?

Goto Sheet 4

Replace A2

Calibrate unit

CC_Prog OK ?

(see Table 3-4)

Yes

Replace A1

Yes

No

No

Yes

Output OK but

meter wrong?

Calibrate current. If still

wrong or will not

calibrate, replace A2

Yes

No

Turn on OCP and

insure Protect trips.

No

Prot trips?

CC_detect* low?

(see Table 3-4)

No

Check cable W9,

replace A1

Replace A2

No

Yes

Yes