Specifications, Table 9: general specifications, Table 10: dc voltage measurement/source – Martel Electronics DMC1410 User Manual
Page 100: Table 11: dc ma measurement/source

10. Specifications
One year at 23°C ± 5°C, unless specified otherwise. Outside of this range
the stability of the measurements is ± 0.003% of FS/°C.
Table 9: General Specifications
Operating Temperature
-10°C to 50°C
Storage Temperature
-20°C to 60°C
Power
4 X AA batteries; Alkaline or optional rechargeable
Battery Life
Approximately 20 hours
Serial Communications ASCII; RS232 or USB depending on the communications
cable used
CE - EMC
EN61326: 2006, ANNEX B Criterion A
Safety
CSA C22.2 No. 1010.1: 1992
Size (with holster)
8.5” x 4” x 2” (21.6cm x 10.2cm x 5.1cm)
Weight
Approximately 28oz. (0.8kg)
Table 10: DC Voltage Measurement/Source
Range
Accuracy
(% of reading ± floor)
Read: Isolated(Upper Display)
0.000V - 30.000V
0.01% ± 2mV
Read: non-Isolated(Lower Display)
0.000V - 20.000V
0.01% ± 2mV
Source
0.000V - 20.000V
0.01% ± 2mV
Maximum current output in voltage ranges is 3mA with an output impedance of <= 1Ω.
Table 11: DC mA Measurement/Source
Range
Accuracy
(% of reading ± floor)
Read: Isolated(Upper Display)
0.000mA - 24.000mA
0.01% ± 2µA
Read: non-Isolated(Lower Display)
0.000mA - 24.000mA
0.01% ± 2µA
Source
0.000mA - 24.000mA
0.01% ± 2µA
Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5V - 30V.