Martel Electronics MC1010 User Manual
Page 42
Range (°C)
Accuracy (°C)
TC Type
Minimum
Maximum
CJC OFF
CJC ON
T
-250.0
-200.0
1.5
1.7
-200.0
0.0
0.5
0.7
0.0
400.0
0.2
0.4
E
-250.0
-200.0
1.0
1.2
-200.0
-100.0
0.3
0.5
-100.0
1000.0
0.2
0.4
R
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.0
1.2
S
1
0.0
200.0
1.7
1.9
200.0
1767.0
1.1
1.3
B
1
600.0
800.0
1.5
1.7
800.0
1000.0
1.2
1.4
1000.0
1820.0
1.0
1.2
C
0.0
1000.0
0.5
0.7
1000.0
2316.0
1.5
1.7
XK
-200.0
800.0
0.2
0.4
BP
0.0
800.0
1.9
2.1
800.0
2500.0
0.6
0.8
L
-200.0
900.0
0.2
0.4
U
-200.0
0.0
0.4
0.6
0.0
600.0
0.2
0.4
N
-200.0
-100.0
0.8
1.0
-100.0
1300.0
0.3
0.5
CJC error outside of 23 ± 5°C is 0.05°C/°C
1. For thermocouple measure mode on B, R and S; round the specification up or down accordingly as there
is no resolution past the decimal point.
Table 17: RTD Read and Source
RTD Type
Range (°C)
Accuracy (°C)
Minimum
Maximum
Measure 4W/Source
Measure 3W
PT385, 10 ohm
-200.0
-80.0
1.3
2.5
-80.0
0.0
1.3
2.5
0.0
100.0
1.4
2.7
100.0
300.0
1.5
3.0
300.0
400.0
1.6
3.1
400.0
630.0
1.8
3.3
630.0
800.0
1.9
3.6
42
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)