beautypg.com

Martel Electronics MC1010 User Manual

Page 42

background image

Range (°C)

Accuracy (°C)

TC Type

Minimum

Maximum

CJC OFF

CJC ON

T

-250.0

-200.0

1.5

1.7

-200.0

0.0

0.5

0.7

0.0

400.0

0.2

0.4

E

-250.0

-200.0

1.0

1.2

-200.0

-100.0

0.3

0.5

-100.0

1000.0

0.2

0.4

R

1

0.0

200.0

1.7

1.9

200.0

1767.0

1.0

1.2

S

1

0.0

200.0

1.7

1.9

200.0

1767.0

1.1

1.3

B

1

600.0

800.0

1.5

1.7

800.0

1000.0

1.2

1.4

1000.0

1820.0

1.0

1.2

C

0.0

1000.0

0.5

0.7

1000.0

2316.0

1.5

1.7

XK

-200.0

800.0

0.2

0.4

BP

0.0

800.0

1.9

2.1

800.0

2500.0

0.6

0.8

L

-200.0

900.0

0.2

0.4

U

-200.0

0.0

0.4

0.6

0.0

600.0

0.2

0.4

N

-200.0

-100.0

0.8

1.0

-100.0

1300.0

0.3

0.5

CJC error outside of 23 ± 5°C is 0.05°C/°C

1. For thermocouple measure mode on B, R and S; round the specification up or down accordingly as there

is no resolution past the decimal point.

Table 17: RTD Read and Source

RTD Type

Range (°C)

Accuracy (°C)

Minimum

Maximum

Measure 4W/Source

Measure 3W

PT385, 10 ohm

-200.0

-80.0

1.3

2.5

-80.0

0.0

1.3

2.5

0.0

100.0

1.4

2.7

100.0

300.0

1.5

3.0

300.0

400.0

1.6

3.1

400.0

630.0

1.8

3.3

630.0

800.0

1.9

3.6

42