PRES_UNIT?
Returns the pressure units. For example if the unit on the display is psi the command will
return:
PSI
PULSE_CNT
Sets the number of pulses the calibrator will produce when it is triggered while in pulse mode.
For example;
PULSE_CNT 3000 will set the number of pulses to 3000.
PULSE_CNT?
Returns the number of pulses in the pulse train. Using the above example, the returned value
would be:
3000
REMOTE
Puts the calibrator in remote mode. From the remote mode the user can still use the keypad
to get back to local unless the command LOCKOUT was entered before REMOTE. Than the
keypad is totally locked out, and the user has to send the LOCAL command to get back to
local operation.
RTD_TYPE
Sets the RTD type. The following is a list of RTD types they way they should be entered after
the command:
PT385_10;
PT385_50;
PT385_100;
PT385_200;
PT385_500;
PT385_1000;
PT392_100;
PTJIS_100;
Ni120;
Cu10; Cu50;
Cu100;
YSI_400;
OHMS;
CUSTOM;
For example:
RTD_TYPE PT385_10 sets RTD type to Pt385-10
RTD_TYPE?
Returns the RTD type.
RTD_WIRE
Sets the number of wires used for connection in measuring RTDs. The calibrator measures
RTDs using 2-, 3-, and 4-wire connections. After the command, enter 2W for 2- wire, 3W for 3-
wire, and 4W for 4-wire. For example:
RTD_WIRE 4W sets the connection to 4-wire
37